Description
Overview
Essential details
Place of Origin:
Guangdong, China
Brand Name:
JCr
Model Number:
JIBPT–01-BD
Disinfecting Type:
EO
Properties:
Medical Materials & Accessories
Size:
233*125*37.6mm
Stock:
Yes
Shelf Life:
3 years
Material:
PVC
Quality Certification:
ce
Instrument classification:
Class III
Safety standard:
ISO 13485
Product name:
Single IBP kit-BD type
Certificate:
CE/IS013485
Supply Ability
Supply Ability
400000 Piece/Pieces per Month
Packaging & delivery
Packaging Details

A case of 40 sets

Port

Shenzhen

Picture Example:
Lead time:
Quantity(pieces) 1 – 1000 1001 – 5000 >5000
Lead time (days) 7 14 To be negotiated
Video Description

00:00

00:45

Product Description
Properties
Medical Materials & Accessories
Product Name
BD disposable IBP transducer
Brand Name
JCR
Model Number
JIBPT-01-BD
Details Images
Main technical parameters and working environment requirements:
Disposable Pressure Transducer
Barometric pressure
70~106Kpa
Relative humidity
10~90% (non-condensing)
Operating pressure range
-50 ~ + 300 mmHg
Sensitivity
5.0μV/V/mmHg ±3%
Nonlinearity and hysteresis
full-scale reading±1.5%
Input impedance
1200Ω ~3200Ω
Zero pressure offset
-20mmHg~+20mmHg
Thermal offset shift
≤±0.3mmHg/°C
Offset drift
After warm-up for 20 seconds, drift within 2 mmHg within 8 hours
Thermal span shift
≤±0.1%/°C
Frequency response
Standard pressure set (48 “/ 12”) is 40 Hz; Separate sensor > 200 Hz
Defibrillator withstand
Defibrillation depends on the final connection of the equipment.
Leakage current
Leakage current is dependent on the final connection of the equipment
Overpressure load
-400~+4000mmHg
Shock resistance
Withstand falling from one meter for three times

Light sensitivity

When exposed to a 3400°K tungsten light, candle light at 3000 feet, it’s less than 1 mmHg under rated voltage.
Storage and transport environment requirements
– 20 ° C~ + 60 ° C, < 90%(non-condensing), no corrosive gas and indoor ventilation, and to avoid high temperature and cold.
Contact with the human body
≤168h
Intra-system velocity
Flow
2-5mL/h
Intra-system velocity refers to the velocity under 6.00 VDC and 25°C, unless there are additional instructions.
Product shelf-life
3 years
Company Introduction
Certifications
Exhibition
Packing&Shipping