DOUBLE CHANNEL SYRINGE PUMP
Model: 500III
Manufacturer: PERLONG MEDICAL EQUIPMENT
FEATURES OF DOUBLE CHANNEL SYRINGE PUMP 500III
Three working modes:rate mode,time-volume mode,dosage-weight mode.
Unique Human Voice Alarm system and accurate injection rate make the injection process safer and more reliable.
Have store many brands syringes,automatically calibration and compatible with any brand syringes
Large LCD screen displays working status.
KVO and Bolus functions.
SPECIFICATIONS OF DOUBLE CHANNEL SYRINGE PUMP 500III
MODEL
500III
Special function
With drug library
Rate
50/60ml Syringe: 0.1ml/h~999.9ml/h (0.1ml/h step)
1,000ml/h~1,500ml/h (1ml/h step)
30ml Syringe: 0.1ml/h~900.0ml/h (0.1ml/h step)
20ml Syringe: 0.1ml/h~600.0ml/h (0.1ml/h step)
10ml Syringe: 0.1ml/h~300.0ml/h (0.1ml/h step)
Flow Rate Accuracy
Within ±3% (after correct calibration)
Mechanical precision
Within ±2%
Bolus Rate
50/60ml Syringe: 1,200ml/h
30 ml Syringe: 720ml/h
20 ml Syringe: 480ml/h
10 ml Syringe: 240ml/h
Purge Rate
50/60ml Syringe: 1,500ml/h
30ml Syringe: 900ml/h
20ml Syringe: 600ml/h
10ml Syringe: 300ml/h
Volume Limit
0.1ml~999.9ml
Total Injection Volume
0.1ml~9999.9ml (0.1ml step)
Occlusion
High: 800mmHg ±200mmHg (106.7kPa±26.7kPa
Medium: 50/600mmHg ±100mmHg (66.7kPa±13.3kPa
Low: 300mmHg ±100mmHg (40.7kPa±13.3kPa )
Alarms
Injection soon finish, end of injection,occlusion, improperly installation of syringe, wrong setting, low battery, syringe loose
etc.
Power Source
AC 100V~240V, 50/60Hz; Internal rechargeable Li battery, capacity≥1,600mAh, 4 hours internal battery backup
KVO rate
1ml/h
Fuse
F1AL/250/60V, 2pcs inside
Power Consumption
30VA
IP Classification
IPX1
Equipment Classification
Class II, internal power supply, Type BF
Operating Condition
Ambient temperature: +5℃~+40℃
Relative humidity
20~90%
Transport & Storage Condition
Ambient temperature: -30℃~+55℃
Relative humidity
≤95%
Dimension
280mm (L) × 210mm (W) × 130mm (H)
Weight
2.2kg (net weight)
PRODUCT SHOW