Brand : Allied
Model : Jupiter Hospital Ventilator
?
Jupiter Hospital Ventilator Specification :-
?
?Time cycled pressure limited electronically controlled ventilator?
?Unique inbuilt calculator ?to guide initial setting of ventilation parameters and inspiratory flow?
?Pneumatically Driven Ventilator
Driven from pipeline/cylinder/ Air compressor
Mains and battery operated.
Bar graph displays dynamic airway pressure showing upper and lower pressure limits.
Settable ?parameters – BPM, I:E ratio pressure limits and inspiratory flow rates
60%? to?100% FiO2 settings.
Built-in audio-visual alarms .
Medical Air Compressor (Optional)
?
1.) Inverse Ratios
The ?ventilator ?has? ?the ?ability to ?adjust? ?substantially? ?when ?clinical conditions require inverse I:E.
2.) Inspiratory flow cut off
?The ?inspiratory ?phase changes to ?expiratory ?in case the ?dynamic pressure reaches the high-pressure limit. High-pressure Audio-visual alarm will also be activated.
3.) Apnea alarm/ low airway pressure alarm
If the dynamic pressure fails to go above ?the low-pressure limit with in??15sec, then the low airway audio visual pressure alarm will be activated.
4.) Works on A-CMV
With the Preset (-) 5 cmH2O of trigger level the ventilator has the ability to synchronize with the patient with the patient breath