Description

AKAS syringe (infusion) pump combines high levels of safety and performance to provide for a basic and reliable Operation. Akash syringe pumpis designed in such a way that it offers several advanced features ideal for general, critical care, neonatal and cancer therapy infusions where accuracy is key. At their simplest, Akash infusion Pump comprises a?user interface that controls the electromechanical pumping mechanism which is responsible for driving fluid from a container (bag, bottle or syringe) through a flexible tubing into a patient?s catheter. Buy with us and get the best akas syringe pump price.?

Key Features of Akash Syringe (infusion) Pump:
Drop/min setting
Ml/hr setting
Infused volume & drops display
Elapsed time display
Target volume & Target time
Setting ? Auto calculation
Built-in rechargeable battery
Excess/Less flow alarm
Less flow alarm due to occlusion
Programmable to any I.V tubing
Suitable for micro / Adult infusions
Audible and visual alarms
Keep vein open(KVO)
?Air in Line? Detector
Bolus Function
Different colour LED?s
LCD Screen with Help Function
Free Flow Protection
Specifications of Akas Syringe infusion pump (Model- Info Max):
PRODUCT NAME
Akas infusion pump
BRAND
Akas
MODEL
INFU MAX
POWER SUPPLY
?AC 220V to 240V 50Hz
BATTERY BACKUP
4hrs @ 8 drops/min or 25 ml/hr (Auto Power Saving Mode)
POWER CONSUMPTION
10VA
PUMPING MECHANISM?
Linear Peristaltic
DROP SENSOR
External ?Air in line?
DETECTOR
Internal
DISPLAY
LCD and 7 Segment LED Display
SAFETY FEATURES
Free flow protection

Excess / Less Alarm

Air in Line Protection

Sensor Malfunction Protection
I.V SET ml CONVERSION
Adult I.V sets ? 12 to 26 drops/ml in steps of 0.5

Micro I.V sets ? 50 to 85 drops/ml in steps of 1
I.V SET TUBING
Accepts all Brands of I.V set
DROP RATE RANGE
1 to 150 drop/min for Adult

1 to 100 drop/min for Micro
VOLUME RATE RANGE
0.5 to 900 ml/hr for Adult

0.1 to 100 ml/hr for Micro
DROP RESOLUTION
1 drop/min
BOLUS FUNCTION
Auto Bolus, Purge Bolus
BOLUS RATE FLOW
10 to 900mL/Hr ? Adult

10 to 100mL/Hr ? Micro
FLOW RATE RESOLUTION?

SEQUENTIAL PRESS
Adult : 0.5 to 100ml/hr Step 0.1ml/hr

100 to 900ml/hr Step 1ml/hr

Micro : 0.1 to 100ml/hr Step 0.1ml/hr
PRESS & HOLD?
?Adult ? 0.5 to 100ml/hr Step 2 ml/hr

100 to 900ml/hr Step 10 ml/hr

Micro : 0.1 to 100ml/hr Step 2 ml/hr

TARGET VOLUME

1 to 990ml

TARGET VOLUME STEPS

Adult:1 to 100ml Step 1 ml
100 to 990ml Step 10ml
Micro : 0.5 to 10ml Step 0.1ml

10 to 100ml Step 1ml

100 to 500ml Step 10ml

BOLUS VOLUME?

0.1 to 50mL

ELAPSED TIME

00:00 to 99:59 Hr

KVO

off to 3dr/min or 3ml/hr steps 1

BATTERY CHARGING PERIOD

16 hrs automatic

BATTERY TYPE

Rechargeable Ni-Cd, Built-in & standard

ALARM INDICATION?

Target volume, Target time, Excess flow, Less flow, Air in line, Battery low, KVO

PURGE FLOW

?Flushing of IV tube possible

MENU-DRIVEN

Self Guided / Help Assisted Interactive Functions , Self Diagnosis of Alarm Causes

HOUSING??

ABS plastic

DIMENSION

140?mm x 140?mm x200mm

WEIGHT

?2.8 kg

CERTIFICATION

CE, ISO 13485:2003, ISO 9001:2008