Product Details:
?
Types
Biphasic Defibrillators
Operation Mode
Semi-Automatic
Brand
Allied
Specifications:
General:
Display:? 7? Blue LCD with LED backlit ,?320 x 240 pixels
Display area: 120mm x 90mm
Dimensions: H180 x W230 x D210mm (max.) without Paddle
Weight: 4.5 Kgs
Defibrillator:
Operating Mode: Manual
Waveform: Biphasic
Energy: 2 to 200 joules in 12 steps
Synchronous: Energy delivery begins within 60 ms of the QRS Peak
Charging Time: Less than 8 seconds
Paddle Assembly: External paddle assembly with coil retractable high voltage cable Adults & Pediatric paddle integrated to the same handle
Check Mode: Manual test mode
Operation: By front Panel keyboard and paddle buttons
ECG:
Frequency Response: 0.5 Hz to 40 Hz (ECG)
Frequency Response: 0.5 Hz to 30 Hz (Paddle)
Patient isolation: Optical isolation
Lead selection: Lead I, II and III
Gain setting: 0.5, 1.0 and 2.0
CMRR: > 90 db
Sweep Speed: 12.5, 25.0, 50.0 mm/sec
Waveform 1: ECG
Waveform 2: 4sec delayed (cascaded)
Data Output: Analog output of ECG
Digital filter: Digital signal processing for good quality of ECG
Freeze mode: To freeze last 4 seconds ECG
Heart Rate:
Range: 30-250 BPM
Accuracy: ? 2 BPM
Resolution: 1 BPM
Tone: Beep on every QRS detected
Low Alarm Limit Range: 30-245 (step of 5)
High Alarm Limit Range: 40-250 (step of 5)
Printer:
Paper Type: Thermal paper 50/55 mm width, 20 meters length
Event Printing
ECG Mode: Prints 5 Seconds pre & post ECG
ECG in Defib
Mode: Prints 5 Seconds pre & post ECG
Defib Mode: Stores & prints 10 Seconds of Defib Shock data i.e. 5 seconds pre-critical & 5 seconds post critical Maximum events that can be store Memory of 10
Manual Defib
Check Mode: Prints the status of Check Mode
Print Annotates: Time, Date, ECG Lead, Mode, Shock Status & selected Energy
Delay: 10 Seconds
Power Supply (AC Mains Through Adaptor):
Input: 150-276VAC, 50Hz, 60Hz, 12V DC for ambulance
Output: 17V@4A
Battery: 12V/3.8 Ah Rechargeable
Backup time: 3 hours min on full charge in ECG mode /50 shocks of 150 Joules in defib mode
Power consumption: 30 Watts (max) in ECG mode
?
Optional Items :
Non-Invasive Pacing
SPO2
NIBP
AED
EtCO2 Monitoring
Environmental Conditions:
Operating Temp: (-)5 ?C ~ (+)50 ?C
Heart Rate Humidity: Up to 95 % non-condensing
?
Additional Information:
Item Code:?Cardiasafe
Delivery Time:?2-4 weeks
Packaging Details:?Domestic/?Export Worthy