Description

Description

DESCRIPTION 

The New E4500-3 industrial gas analyzer measures Oxygen (O2), Carbon Monoxide (CO), Nitric Oxide (NO) and Nitrogen Oxide (NOx) gas sensors calculation and Carbon Dioxide (CO2). Display and print combustion efficiency, along with ALL of the necessary temperature, draft, and pressure measurements. The E4500-3 is a commercial and industrial gas analyzer that is a valuable tool designed for emissions monitoring and maintenance and tuning of combustion processes including boilers, burners, gas and diesel engines, turbines, furnaces, kilns, heaters, and laboratory analysis.
Features

Measure Oxygen (O2), Carbon Monoxide (CO) and Carbon Dioxide (CO2), Nitric Oxide (NO) and Nitrogen Oxides (NOx)
Full Color Graphic Display Screen
Built-In Impact Printer (non-fading paper)
Internal Memory (up to 2000 tests)
Data Logging with time and date stamp
PC Software Package with USB & Bluetooth Communications
Dilution Pump For CO Auto-Range Measurements Up To 50,000 ppm
Low NOx Capable with 0.1ppm resolution & high accuracy
Draft and Differential Pressure Measurements
10 Selectable Fuel/Oil Types Including Bio-Heat 5%
Efficiency, Excess Air, and CO2 Calculations
Stack Gas and Air Temperature Measurements
External Water Trap for Condensate Removal
Rechargeable Lithium Ion Battery Pack
Unbreakable Metal Hose Connectors
ALL Gas Sensors are Pre-Calibrated and Field Replaceable

E4500-3 Combustion Analyzer Measurement Specifications

Parameter
Sensor
Range
Resolutoin
Accuracy

Oxygen (02)
Electrochemical
0 to 25%
0.1%
±0.2% volume

Carbon Monoxide (CO)
H² Compensated
WITH Built-in NOx Filter
Electrochemical
0 to 8,000 ppm
1 ppm
±±10 ppm (0 to 200ppm)
±5% rdg (201 to 2000ppm)
±10% rdg (2001 to 8000ppm)

Carbon Monoxide (CO) DILUTED
Electrochemical
0.8% – 5.00%
0.01%
±10% rdg

Carbon Dioxide (CO2)
Calculated
0 to 99.9%
0.1%

Nitric Oxide (NO)
Electrochemical
0 to 5000 ppm
1 ppm
±5 ppm (0-100ppm)
±5% rdg (101-5000ppm)

NOX Measurement
Calculated
0 to 5000 ppm
1 ppm

Low NOx
Electrochemical
0 to 500 ppm
0.1 ppm
±2 ppm (0.0-40.0ppm)
±5% rdg (40.1-500.0ppm)

Temperature Air
Pt100
-10 to 212°F (-10 to 100°C)
0.1°F or °C
±1°F/°C

Temperature gas
Type K Thermocouple
-4 to 2280°F (-20°C to 1250°C)
0.1°F or 0.1°C
± 2°F

Differential Temperature
Calculated
-10 to 1830°F (-23°C to 999°C)
1°F or °C
±0.5 °C (-20 to 100 °C)
±0.5% rdg (101 to 1250°C)

Pressure/Draft
Semiconductor
±0 to 80 inH²O
0.001 inH2O
±1.0% rdg

Combustion Efficiency
Calculated
0 to 100%
0.1%

Excess Air
Calculated
0 to 850%
1%

E4500-3 General Specifications

Display
Graphic 272 x 480 pixels, backlit, colour TFT 4.3″

Printer
Thermal integrated with easy loading paper and sensor for the presence of paper

Printout heading
4 lines x 24 characters, customisable by the user

Printer power supply
From the analyzer battery pack

Internal data storage
2000 complete data analyses including date, time and client’s name

Communications port
USB with mini-USB connector

Bluetooth Communication
Communication range: <100 meters (free field) Power Supply Rechargeable battery, Li-Ion battery pack with internal protection circuit Battery life 12 hours of continuous operation (printing excluded) Charging time 5 hours to recharge from 0% to 90% Operating Temperature 23 to 104°F (-5°C to +40°C) Storage Temperature 14 to 122°F (-10°C to +50°C) Operating humidity range 20% to 80% RH Air pressure Atmospheric Outer dimensions analyzer 31 x 9 x 6 cm (H x W x D) Weight 1 kg