Buy Fluke 1503 Insulation Resistance Tester Meter from safety Nigeria, your online supplier of flukes with fast delivery. 
Key features

A rugged, compact insulation resistance tester for residential and commercial electricians

Performs insulation tests from 0.01 MΩ to 2000 MΩ

Offers two insulation test voltages: 500 V and 1000 V

Includes live circuit detection to prevent insulation test if voltage >30 V is detected

Measures resistance from 0.01 Ω to 20.00 kΩ

Product overview: Fluke 1503 Insulation Resistance Meter

Fluke 1503 Insulation Resistance Tester handles all the basic insulation testing needs
The Fluke 1503 is a great choice for basic residential and commercial electrical insulation testing. This rugged, compact tool fits in the palm of your hand. The 1503 is the same form factor as the Fluke 1507 but with a few less features. You can count on it to handle the most common tests at a most affordable price. Additional features, like the remote probe, reduce the time needed to perform repetitive testing.
Other useful capabilities:

Measures AC/DC voltage from 0.1 V to 600 V

Includes auto-discharge of capacitive voltage for added user protection

Provides lo-ohms earth-bond continuity (200 mA)

Simplifies repetitive or hard-to-reach testing with the remote test probe

Saves battery power with auto power off

Presents results on a large, backlit display

Features CAT IV 600 V overvoltage category rating

Comes with remote probe, test leads, test probes, alligator clips, and one-year warranty

Accepts optional Fluke TPAK magnetic hanging system to free your hands for other work

Runs on four AA alkaline batteries (NEDA 15 A or IEC LR6) for at least 1000 insulation tests

Specifications: Fluke 1503 Insulation Resistance Meter

AC/DC Voltage Measurement

Accuracy

Range

600.0 V

Resolution

0.1 V

50 Hz to 400 Hz ±(% of Rdg + Digits)

±(2% + 3)

Input impedance

3 MΩ (nominal), < 100 pF Common mode rejection ratio (1 kΩ unbalanced) > 60 dB at DC, 50 or 60 Hz

Overload protection

600 V rms or DC

Earth Bond Resistance Measurement

Range/Resolution

20.00 Ω

0.01 Ω

200.0 Ω

0.1 Ω

2000 Ω

1.0 Ω

20.00 kΩ

0.01 kΩ

Accuracy

±(1.5% + 3)

Overload protection

2 V rms or DC

Open circuit test voltage

> 4.0 V, < 8 V Short circuit current > 200.0 mA

Insulation Specifications

Measurement range

0.01 MΩ to 2000 MΩ

Test voltage

500 V, 1000 V

Test voltage accuracy

+ 20%, – 0%

Short circuit current

1 mA nominal

Auto discharge

Discharge time < 0.5 second for C = 1 µF or less Live circuit indicator Inhibit test if terminal voltage > 30 V prior to initialization of test

Maximum capacitive load

Operable with up to 1 µF load

Measure accuracy

500 V

±(1.5% + 5)

1000 V

±(1.5% + 5) to 2000 MΩ, ±(10% + 3) above 2000 MΩ

General Specifications

Maximum voltage applied to any terminal:

600 V AC rms or DC

Storage temperature

-40°C to 60°C (-40°F to 140°F)

Operating temperature

-20°C to 55°C (-4°F to 131°F)

Temperature coefficient

0.05 x (specified accuracy) per °C for temperatures < 18°C or > 28°C (< 64°F or > 82°F)

Relative humidity

0% to 95% @ 10°C to 30°C (50°F to 86°F)

0% to 75% @ 30°C to 40°C (86°F to 104°F)

0% to 40% @ 40°C to 55°C (104°F to 131°F)

Vibration

Random, 2 g, 5-500 Hz per MIL-PRF-28800F, Class 2 instrument

Shock

1 meter drop per IEC 61010-1 2nd Edition (1 meter drop test, six sides, oak floor)

Electromagnetic compatibility

In an RF field of 3 V/M, accuracy = specified accuracy (EN 61326-1:1997)

Safety

Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 NO. 61010-1-04, and IEC/EN 61010-1 2nd Edition for measurement category IV 600 V (CAT IV)

Certifications

CSA per standard CSA/CAN C22.2 No. 61010.1-04; TUV per standard IEC/EN 61010-1 2nd Edition

Batteries

Four AA batteries (NEDA 15A or IEC LR6)

Battery llife

Insulation test use:

Tester can perform at least 1000 insulation tests with fresh alkaline batteries at room temperature. These are standard tests of 1000 V into 1 MΩ with a duty cycle of 5 seconds on and 25 seconds off.

Resistance measurements:

Tester can perform at least 2500 earth bond resistance measurements with fresh alkaline batteries at room temperature. These are standard tests of 1 Ω with a duty cycle of 5 seconds on and 25 seconds off.

Size (H x W x L)

5.0 x 10.0 x 20.3 cm (1.97 x 3.94 x 8.00 in)

Weight

550 g (1.2 lb)

IP rating

IP40

Altitude

Operating

2000 m CAT IV 600 V, 3000 m CAT III 600 V

Non-operating (storage)

12,000 m

Over-range capability

110% of range

Included accessories

TL224 Test Leads, TP74 Test Probes, clips PN 1958654 (red) and PN 1958646 (black), holster and remote probe

Reference Specific ReferencesCondition New product Andrew U03/20/2020Fluke 1503 Insulation Resistance Tester Metertrue resistance testerWrite your reviewFluke 1503 Insulation Resistance Tester MeterQuality12345Delivery12345Communication12345Pricing12345Title for your review*Your review*Required fields* Send or Cancel

Reviews
Ratings

0.0

0 Product Ratings
5
0
4
0
3
0
2
0
1
0

Review this product

Share your thoughts with other customers

Write a review

Reviews

There are no reviews yet.

Fluke 1503 Insulation Resistance Tester Meter 1,000.00
Home Shop 2 Cart Account
Shopping Cart (2)