Specification of?GE Carescape R860 Ventilator:
BrandGEModel Name/NumberCarescape R860Patient Age GroupAdults, Pediatric Patients, NeonatalTidal VolumeAdult: 100 to 2000 ml, Pediatric: 20 to 300 mlNeonatal: 2 to 50 mlVentilation ModesA/C VC, A/C PC, A/C PRVC, SIMV VC, SIMV PC, CPAP/PS, SBT, APRV, BiLevel, BiLevel VG, SIMV PRVC, VS, NIV, nCPAPRespiratory RateNeonatal: 3 to 150 bpm for A/C VC, A/C PC, A/C PRVC; 2 to 60 bpm for SIMV VC, SIMV PC, SIMV PRVC; Off, 1 to 30 bpm for nCPAP
Adult/Pediatric: 3 to 120 bpm for A/C VC, A/C PC, A/C PRVC; 1 to 60 bpm for SIMV VC, SIMV PC, SIMV PRVCControls
TriggerFlow trigger:- Neonatal: 0.2 to 1 L/min (increments of 0.05 L/min), 1 to 3 L/min (increments of 0.1 L/min), 3 to 9 L/min (increments of 0.5 L/min); Adult/Pediatric: 1 to 3 L/min (increments of 0.1 L/min), 3 to 9 L/min (increments of 0.5 L/min)
Pressure trigger:- Adult/Pediatric: ?10 to ?3 cm H?O (increments of 0.5 cm H?O), ?3 to ?0.25 cm H?O (increments of 0.25 cm H?O)Pressure Support0 to 60 cm H?O for SIMV VC, SIMV PC,SIMV PRVC, BiLevel, BiLevel VG and CPAP/PS (increments of 1 cm H?O), 0 to 30 cm H?O for NIV(increments of 1 cm H?O)FlowNeonatal: 0.2 to 30 L/min, Pediatric: 2 to 72 L/min, Adult: 2 to 160 L/minMaximum Peak FlowAdult: 208 L/min; Pediatric: 100 L/min; Neonatal: 30 L/minInspiratory/ Expiratory RatioNeonatal: 1:199 to 40:1 in BiLevel and APRV, Adult/Pediatric: 1:79 to 60:1 in BiLevel and APRVTrigger Window80% of expiration timeBias Flow RateNeonatal: 2 to 15 L/min (increments of 0.5 L/min) for nCPAP; Adult/Pediatric: 2 to 10 L/min (increments of 0.5 L/min), 8 to 20 L/min for NIV (increments of 0.5 L/min)Settings
Inspiratory PressureNeonatal: 1 to 25 cm H?O (increments of 1 cm H?O) in nCPAP; Adult/Pediatric: 1 to 98 cm H?O (increments of 1 cm H?O)PEEP/CPAPPEEP:- Neonatal: 2-15 cm H?O (increments of 1 cm H?O) in nCPAP
Adult/Pediatric: Off, 1 to 50 cm H?O (increments of 1 cm H?O), 2-20 cm H?O (increments of 1 cm H?O) in NIVInspiratory timeNeonatal: 0.1 to 10 sec, 0.1 to 2 sec in nCPAP, 0.1 to 1 sec (increments of 0.01), 1 to 4 sec (increments of 0.1), 4 to 10 sec (increments of 0.25)
Adult/Pediatric: 0.25 to 15 sec, 0.25 to 5 sec in NIV, 0.25 to 1 sec (increments of 0.05), 1 to 4 sec (increments of 0.1), 4 to 15 sec (increments of 0.25)Expiratory timeNeonatal: 0.25 to 29.9 sec Invasive vent modes, Adult/Pediatric: 0.25 to 59.75 secMonitoring
Peak Inspiratory Pressure7 to 100 cm H?O (increments of 1 cm H?O), 9 to 100 cm H?O (increments of 1 cm H?O)in NIV and nCPAPAmbient Conditions
Temperature (device) During Operation10? to 40?CRelative humidity During Operation15 to 95% RH Non-condensingAltitudeOperating range: ?400 to 3000 m/525 to 800 mmHg , Storage range: ?400 to 5860 m/375 to 800 mmHgTemperature During Storage & Transportation?20? to 60?CRelative humidity During Storage & Transportation15 to 75% RH Non-condensingPerformance
Intermittent PEEPHigh: 1 to 20 cm H2O, OffMinute volumeLow: 0.01 to 40 L/min, High: 0.02 to 99 L/minOperating Data
Power Consumption< 200 VADimensions151.77 x 38 x 36 cmWeight31 kg (not including cart), 78 kg (including cart)Power Supply
Battery TypeLead-acid gelAlarm
Inspiratory O? concentration (FiO?)Low: 18 to 99%, High: 24 to 100%, OffAirway Pressure?20 to +120 cm H?OApnea alarm time5 to 60 secondsAudible Alarm Volume Adjustable0, 10, 20, and 30 sec, OffHigh Inspiratory Pressure1 to 98 cm H?O (increments of 1 cm H?O)Low Inspiratory PressureOff, 1 to 50 cm H?O (increments of 1 cm H?O)High Respiratory Rate2 to 150/min, OffCircuit Leak10 to 90%, OffEscalating AlarmsHigh priority alarms escalate to a higher pitch if unattended for a specified timeAuto LimitsAlarm limits calculated on the current measured values for selected parametersDisplay Motion
Vertical Tilt162.1? in the raised position, 47.6? in the lowered positionHeight Adjustment24.95 cm / 9.8 inchAdditional Details
Inspiratory Pressure Limit7 to 100 cm H?O for A/C VC and SIMV VC(increments of 1 cm H?O)CO? production (VCO?)20 to 1000 mL/minRise time0 to 500 ms of the inspiratory period for pressure supported breaths only. Active in SIMV VC, SIMV PC, SIMV PRVC, CPAP/PS, BiLevel VG, BiLevel, and VS (increments of 50 ms)Features of?GE Carescape R860 Ventilator:
Lung Protection StrategyInnovative user interfaceMenu-free navigationMeasuring Functional Residual CapacityApplications?GE Carescape R860 Ventilator:
Emergency Care & ICUProduct Description of?GE Carescape R860 Ventilator:
Ventilator features expert user tools, an innovative user interface, and is inherently familiar the first time you use it.
Featuring expert user tools and an innovative user interface, the Carescape R860 is inherently familiar the first time you use it.
Navigating the past, present, and future states of your ventilated patients is the foundation of the Carescape R860 user interface. From the center workspace view that highlights current patient status, you can swipe left to right to navigate from historical trends on the left (past) to Clinical Decision Support tools on the right (future). Your settings are recalled even as you navigate across workspaces.
By replacing menus with workspace views, the Carescape R860 has flattened the learning curve and unlocked ventilation capabilities by making them more accessible and easier to use.